Agilent Flow Cells for HPLC

Agilent provides a comprehensive range of high quality VWD and DAD/MWD flow cells for LC systems to allow best performance for all applications.

The specific repair kit for each flow cell includes all relevant supplies required to produce the highest level of performance in all VWD and DAD/MWDs.

Part Number Product Description Price Qty
product image

Features

Cleaning or Replacing DAD/MWD Flow Cells

  • A decrease in detector performance or unusual noise levels may mean you have dirty flow cell windows
  • Clean and reassemble one side of the flow cell before beginning the other side to prevent mixing the front and rear gaskets, which have different hole diameters
  • While cleaning or replacing flow cell windows, if the washers fall out of the window assembly, they must be inserted in the correct order with a PTFE ring to prevent any leaks from the flow cell window
  • Clean the cell body with water or isopropanol
  • After opening the cell you should always use a new gasket

Specifications

VWD Flow Cell Selection

Typical Column Length Typical Peak Width Recommended Flow Cell
< = 5 cm 0.025 Micro Flow Cell       High Pressure Flow Cell
10 cm 0.05   Semimicro Flow Cell    
20 cm 0.1     Standard Flow Cell
>= 40 cm 0.2    
Typical Flow Rate   0.05 - 0.2 mL/min 0.2 - 0.4 mL/min 0.4 - 0.8 mL/min 1 - 2 mL/min  0.05 - 5 mL/min
Internal Column Diameter   1 mm 2.1 mm 3 mm 4.6 mm  

 

DAD/MWD Flow Cell Selection

Typical Column Length Typical Peak Width Recommended Flow Cell
< = 5 cm 0.025 80/500 nL Flow Cell       High Pressure Flow Cell
10 cm 0.05   Semimicro Flow Cell    
20 cm 0.1     Standard Flow Cell
>= 40 cm 0.2    
Typical Flow Rate   0.05 - 0.2 mL/min 0.2 - 0.4 mL/min 0.4 - 0.8 mL/min 1 - 2 mL/min  0.05 - 5 mL/min
Internal Column Diameter   0.3 - 1 mm 2.1 mm 3 mm 4.6 mm  