Agilent Flow Cells for HPLC

All pictures shown are for illustration purpose only. Actual product may vary.
Agilent provides a comprehensive range of high quality VWD and DAD/MWD flow cells for LC systems to allow best performance for all applications.
The specific repair kit for each flow cell includes all relevant supplies required to produce the highest level of performance in all VWD and DAD/MWDs.
Features
Cleaning or Replacing DAD/MWD Flow Cells
- A decrease in detector performance or unusual noise levels may mean you have dirty flow cell windows
- Clean and reassemble one side of the flow cell before beginning the other side to prevent mixing the front and rear gaskets, which have different hole diameters
- While cleaning or replacing flow cell windows, if the washers fall out of the window assembly, they must be inserted in the correct order with a PTFE ring to prevent any leaks from the flow cell window
- Clean the cell body with water or isopropanol
- After opening the cell you should always use a new gasket
Specifications
-
VWD Flow Cell Selection
Typical Column Length Typical Peak Width Recommended Flow Cell < = 5 cm 0.025 Micro Flow Cell High Pressure Flow Cell 10 cm 0.05 Semimicro Flow Cell 20 cm 0.1 Standard Flow Cell >= 40 cm 0.2 Typical Flow Rate 0.05 - 0.2 mL/min 0.2 - 0.4 mL/min 0.4 - 0.8 mL/min 1 - 2 mL/min 0.05 - 5 mL/min Internal Column Diameter 1 mm 2.1 mm 3 mm 4.6 mm DAD/MWD Flow Cell Selection
Typical Column Length Typical Peak Width Recommended Flow Cell < = 5 cm 0.025 80/500 nL Flow Cell High Pressure Flow Cell 10 cm 0.05 Semimicro Flow Cell 20 cm 0.1 Standard Flow Cell >= 40 cm 0.2 Typical Flow Rate 0.05 - 0.2 mL/min 0.2 - 0.4 mL/min 0.4 - 0.8 mL/min 1 - 2 mL/min 0.05 - 5 mL/min Internal Column Diameter 0.3 - 1 mm 2.1 mm 3 mm 4.6 mm